The implementation of a rapid sample preparation method for the detection of SARS-CoV-2 in a diagnostic laboratory in South Africa
Master Thesis
2022
Permanent link to this Item
Authors
Supervisors
Journal Title
Link to Journal
Journal ISSN
Volume Title
Publisher
Publisher
Faculty
License
Series
Abstract
The SARS-CoV-2 pandemic has resulted in shortages of both critical reagents for nucleic acid purification and highly trained staff as supply chains are strained by high demand, public health measures and frequent quarantining and isolation of staff. This created the need for alternate workflows with limited reliance on specialised reagents, equipment and staff. We present here the validation and implementation of such a workflow for preparing samples for downstream SARS-CoV-2 RT-PCR using liquid handling robots. The rapid sample preparation and inactivation technique evaluated, which included sample centrifugation and heating prior to RT-PCR, showed a 97.37% (95% CI: 92.55-99.28%) positive percent agreement and 97.30% (95% CI: 90.67-99.52%) negative percent agreement compared to nucleic acid purification-based testing. A total of 195 samples were tested as part of the validation. This method was subsequently adopted as the primary sample preparation method in the Groote Schuur Hospital Virology Diagnostic Laboratory in Cape Town, South Africa.
Description
Keywords
Reference:
Marais, G.J.K. 2022. The implementation of a rapid sample preparation method for the detection of SARS-CoV-2 in a diagnostic laboratory in South Africa. . ,Faculty of Health Sciences ,Department of Clinical Laboratory Sciences. http://hdl.handle.net/11427/37618