Development of a patient-specific finite element model of the transcatheter aortic valve implantation (TAVI) procedure

dc.contributor.advisorReddy, B Dayaen_ZA
dc.contributor.authorShirzadi, Mohammad Mehdien_ZA
dc.date.accessioned2017-01-23T07:48:12Z
dc.date.available2017-01-23T07:48:12Z
dc.date.issued2016en_ZA
dc.description.abstractTranscatheter Aortic Valve Implantation (TAVI) is a procedure developed for replacing the defective aortic valve of a patient as an alternative to open heart Surgical Aortic Valve Replacement (SAVR). In the TAVI procedure a prosthetic valve, which is assembled on to a stent, is crimped and delivered to the patient's aortic root site through several available percutaneous means. The percutaneous nature of TAVI, which is its core advantage in comparison to other SAVR procedures, can however also be its main disadvantage. This is due to lack of direct access to the calcified leaflets, and hence reliance on the host tissue for the proper positioning and anchorage of the deployed prosthetic valve. Therefore, it is desired to have a preoperative quantitative understanding of patient-specific biomechanical interaction of the stent and the native valve to be able to maximise the chance of success of the procedure. The aim of this study was to develop a patient-specific Finite Element (FE) model of the Transcatheter Aortic Valve Implantation (TAVI) procedure for two patients, using a model of the 23 mm percutaneous prosthetic aortic valve developed by Strait Access Technologies (SAT), for the purpose of its post-operative performance. In this regard, the image processing software ScanIP was used to extract the 3D models of the patient-specific aortic roots and leaflets from the provided Multi-Slice Computer Tomography (MSCT) images of the patients. An anisotropic hyperelastic material model was implemented for the roots and leaflets, using two and one families of collagen fibres for their tissues respectively. The stent is made of a cobalt-chromium alloy and its mechanical response was modelled as an isotropic elastoplastic material, with a linear elastic initial response, followed by plastic behaviour with isotropic hardening. The prosthetic leaflets are made of polymer and were modelled as an isotropic hyperelastic material, using the provided experimental test data. The results for the first patient showed that the stent maintained its structural integrity after deployment, and successfully pushed the native leaflets back to keep the aortic root clear of all impediments. No obstruction of the coronary ostia was observed, and prosthetic leaflets were seen to function normally. The stent radial recoil was calculated to be between 2 to 4.28 % after deployments. Its foreshortening was calculated to be approximately 20%. The stent was observed to move back and forth by approximately 3 mm in the last simulation step in which cardiac cycle pressure were applied to the aortic root and prosthetic leaflets. Also, two openings were observed between the stent and aortic root wall during this simulation step, which indicates the possibility of paravalvular leakage. From the second patient simulation, it was observed that the 23 mm stent was not a good choice for this patient, and will cause severe damage or tissue tearing. The maximum principal stress in the aortic root and valve tissues were observed to follow approximately the defined collagen fibre directions.en_ZA
dc.identifier.apacitationShirzadi, M. M. (2016). <i>Development of a patient-specific finite element model of the transcatheter aortic valve implantation (TAVI) procedure</i>. (Thesis). University of Cape Town ,Faculty of Engineering & the Built Environment ,Department of Mechanical Engineering. Retrieved from http://hdl.handle.net/11427/22893en_ZA
dc.identifier.chicagocitationShirzadi, Mohammad Mehdi. <i>"Development of a patient-specific finite element model of the transcatheter aortic valve implantation (TAVI) procedure."</i> Thesis., University of Cape Town ,Faculty of Engineering & the Built Environment ,Department of Mechanical Engineering, 2016. http://hdl.handle.net/11427/22893en_ZA
dc.identifier.citationShirzadi, M. 2016. Development of a patient-specific finite element model of the transcatheter aortic valve implantation (TAVI) procedure. University of Cape Town.en_ZA
dc.identifier.ris TY - Thesis / Dissertation AU - Shirzadi, Mohammad Mehdi AB - Transcatheter Aortic Valve Implantation (TAVI) is a procedure developed for replacing the defective aortic valve of a patient as an alternative to open heart Surgical Aortic Valve Replacement (SAVR). In the TAVI procedure a prosthetic valve, which is assembled on to a stent, is crimped and delivered to the patient's aortic root site through several available percutaneous means. The percutaneous nature of TAVI, which is its core advantage in comparison to other SAVR procedures, can however also be its main disadvantage. This is due to lack of direct access to the calcified leaflets, and hence reliance on the host tissue for the proper positioning and anchorage of the deployed prosthetic valve. Therefore, it is desired to have a preoperative quantitative understanding of patient-specific biomechanical interaction of the stent and the native valve to be able to maximise the chance of success of the procedure. The aim of this study was to develop a patient-specific Finite Element (FE) model of the Transcatheter Aortic Valve Implantation (TAVI) procedure for two patients, using a model of the 23 mm percutaneous prosthetic aortic valve developed by Strait Access Technologies (SAT), for the purpose of its post-operative performance. In this regard, the image processing software ScanIP was used to extract the 3D models of the patient-specific aortic roots and leaflets from the provided Multi-Slice Computer Tomography (MSCT) images of the patients. An anisotropic hyperelastic material model was implemented for the roots and leaflets, using two and one families of collagen fibres for their tissues respectively. The stent is made of a cobalt-chromium alloy and its mechanical response was modelled as an isotropic elastoplastic material, with a linear elastic initial response, followed by plastic behaviour with isotropic hardening. The prosthetic leaflets are made of polymer and were modelled as an isotropic hyperelastic material, using the provided experimental test data. The results for the first patient showed that the stent maintained its structural integrity after deployment, and successfully pushed the native leaflets back to keep the aortic root clear of all impediments. No obstruction of the coronary ostia was observed, and prosthetic leaflets were seen to function normally. The stent radial recoil was calculated to be between 2 to 4.28 % after deployments. Its foreshortening was calculated to be approximately 20%. The stent was observed to move back and forth by approximately 3 mm in the last simulation step in which cardiac cycle pressure were applied to the aortic root and prosthetic leaflets. Also, two openings were observed between the stent and aortic root wall during this simulation step, which indicates the possibility of paravalvular leakage. From the second patient simulation, it was observed that the 23 mm stent was not a good choice for this patient, and will cause severe damage or tissue tearing. The maximum principal stress in the aortic root and valve tissues were observed to follow approximately the defined collagen fibre directions. DA - 2016 DB - OpenUCT DP - University of Cape Town LK - https://open.uct.ac.za PB - University of Cape Town PY - 2016 T1 - Development of a patient-specific finite element model of the transcatheter aortic valve implantation (TAVI) procedure TI - Development of a patient-specific finite element model of the transcatheter aortic valve implantation (TAVI) procedure UR - http://hdl.handle.net/11427/22893 ER - en_ZA
dc.identifier.urihttp://hdl.handle.net/11427/22893
dc.identifier.vancouvercitationShirzadi MM. Development of a patient-specific finite element model of the transcatheter aortic valve implantation (TAVI) procedure. [Thesis]. University of Cape Town ,Faculty of Engineering & the Built Environment ,Department of Mechanical Engineering, 2016 [cited yyyy month dd]. Available from: http://hdl.handle.net/11427/22893en_ZA
dc.language.isoengen_ZA
dc.publisher.departmentDepartment of Mechanical Engineeringen_ZA
dc.publisher.facultyFaculty of Engineering and the Built Environment
dc.publisher.institutionUniversity of Cape Town
dc.subject.otherMechanical Engineeringen_ZA
dc.titleDevelopment of a patient-specific finite element model of the transcatheter aortic valve implantation (TAVI) procedureen_ZA
dc.typeMaster Thesis
dc.type.qualificationlevelMasters
dc.type.qualificationnameMSc (Eng)en_ZA
uct.type.filetypeText
uct.type.filetypeImage
uct.type.publicationResearchen_ZA
uct.type.resourceThesisen_ZA
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
thesis_ebe_2016_shirzadi_mohammad_mehdi (1).doc.pdf
Size:
3.71 MB
Format:
Adobe Portable Document Format
Description:
Collections