dc.description.abstractBackground External beam radiotherapy (EBRT) is an important part of the treatment for prostate cancer. To account for organ motion and set-up error, margins are used in radiotherapy planning. These are essential for the delivery of safe and effective treatment. Aim and Setting To measure prostate motion during the course of EBRT on the HalcyonTM at Groote Schuur Hospital. Using published margin recipes, the minimum clinical target volume (CTV)-planning target volume (PTV) expansion margin was calculated. Patients and Methods Prostate motion was evaluated by comparing prostate position on the planning CT to cone-beam CT (CBCT) scans. Prostate position error in the medio-lateral (ML), antero-posterior (AP) and superoinferior (SI) directions was measured for each CBCT. The systematic (Σ) and random (σ) error of prostate motion was calculated. The minimum CTV-PTV margin was determined. Results The mean position error of the prostate was -0.04mm (95% CI -0.14, 0.10), -0.30mm (95% CI -0.70, 0.15) and -0.40mm (95% CI -0.89, 0.00) in the ML, AP and SI directions, respectively. Using Van Herk's margin formula (2.5Σ +0.7σ), the following minimum CTV-PTV margins for the ML, AP and SI directions were calculated: 1.9mm, 6.9mm and 7.8mm, respectively. Using Stroom's margin formula (2Σ +0.7σ), the following minimum CTV-PTV margins for the ML, AP and SI directions were calculated: 1.6mm, 5.8mm and 6.6mm, respectively. Conclusion Based on prostate motion, our institution's current margins are sufficient. However, further studies are necessary to measure other factors that influence the CTV-PTV margin before considering reducing this margin.
dc.identifier.apacitationAlgar, M. (2023). <i>Evaluation of target motion and determination optional treatment margins for prostate cancer treated with external beam radiotherapy on Halcyon</i>. (). ,Faculty of Health Sciences ,Division of Radiology. Retrieved from http://hdl.handle.net/11427/39261en_ZA
dc.identifier.chicagocitationAlgar, Marion. <i>"Evaluation of target motion and determination optional treatment margins for prostate cancer treated with external beam radiotherapy on Halcyon."</i> ., ,Faculty of Health Sciences ,Division of Radiology, 2023. http://hdl.handle.net/11427/39261en_ZA
dc.identifier.citationAlgar, M. 2023. Evaluation of target motion and determination optional treatment margins for prostate cancer treated with external beam radiotherapy on Halcyon. . ,Faculty of Health Sciences ,Division of Radiology. http://hdl.handle.net/11427/39261en_ZA
dc.identifier.ris TY - Thesis / Dissertation AU - Algar, Marion AB - Background External beam radiotherapy (EBRT) is an important part of the treatment for prostate cancer. To account for organ motion and set-up error, margins are used in radiotherapy planning. These are essential for the delivery of safe and effective treatment. Aim and Setting To measure prostate motion during the course of EBRT on the HalcyonTM at Groote Schuur Hospital. Using published margin recipes, the minimum clinical target volume (CTV)-planning target volume (PTV) expansion margin was calculated. Patients and Methods Prostate motion was evaluated by comparing prostate position on the planning CT to cone-beam CT (CBCT) scans. Prostate position error in the medio-lateral (ML), antero-posterior (AP) and superoinferior (SI) directions was measured for each CBCT. The systematic (Σ) and random (σ) error of prostate motion was calculated. The minimum CTV-PTV margin was determined. Results The mean position error of the prostate was -0.04mm (95% CI -0.14, 0.10), -0.30mm (95% CI -0.70, 0.15) and -0.40mm (95% CI -0.89, 0.00) in the ML, AP and SI directions, respectively. Using Van Herk's margin formula (2.5Σ +0.7σ), the following minimum CTV-PTV margins for the ML, AP and SI directions were calculated: 1.9mm, 6.9mm and 7.8mm, respectively. Using Stroom's margin formula (2Σ +0.7σ), the following minimum CTV-PTV margins for the ML, AP and SI directions were calculated: 1.6mm, 5.8mm and 6.6mm, respectively. Conclusion Based on prostate motion, our institution's current margins are sufficient. However, further studies are necessary to measure other factors that influence the CTV-PTV margin before considering reducing this margin. DA - 2023 DB - OpenUCT DP - University of Cape Town KW - Radiation Oncology LK - https://open.uct.ac.za PY - 2023 T1 - ETD: Evaluation of target motion and determination optional treatment margins for prostate cancer treated with external beam radiotherapy on Halcyon TI - ETD: Evaluation of target motion and determination optional treatment margins for prostate cancer treated with external beam radiotherapy on Halcyon UR - http://hdl.handle.net/11427/39261 ER - en_ZA
dc.identifier.urihttp://hdl.handle.net/11427/39261
dc.identifier.vancouvercitationAlgar M. Evaluation of target motion and determination optional treatment margins for prostate cancer treated with external beam radiotherapy on Halcyon. []. ,Faculty of Health Sciences ,Division of Radiology, 2023 [cited yyyy month dd]. Available from: http://hdl.handle.net/11427/39261en_ZA
dc.language.rfc3066eng
dc.publisher.departmentDivision of Radiology
dc.publisher.facultyFaculty of Health Sciences
dc.subjectRadiation Oncology
dc.titleEvaluation of target motion and determination optional treatment margins for prostate cancer treated with external beam radiotherapy on Halcyon
dc.typeThesis / Dissertation
dc.type.qualificationlevelMasters
dc.type.qualificationlevelMMed
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
thesis_hsf_2023_algar marion.pdf
Size:
1.86 MB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.72 KB
Format:
Item-specific license agreed upon to submission
Description:
Collections