Reducing radiation dose for a linear slot scanning digital X-ray machine using a filtration technique

dc.contributor.authorPerks, Timothy Daviden_ZA
dc.date.accessioned2015-09-15T10:23:35Z
dc.date.available2015-09-15T10:23:35Z
dc.date.issued2014en_ZA
dc.description.abstractThis study describes the development of a filtration technique applied to the Lodox Statscan linear slotscanning digital X-ray system to reduce radiation dose to paediatric patients whilst preserving diagnostic image quality. The Statscan is an FDA approved, commercially available digital X-ray system commonly used for trauma and emergency patients. The Statscan provides significantly lower radiation dose to patients than conventional Xray systems for comparable studies without loss of image quality. This is particularly beneficial in paediatric radiology, where the risks associated with ionizing radiation are much higher. A static dose prediction model for the Statscan which was previously developed at the University of Cape Town has been adapted to create a dynamic dose prediction model which allows the user to adjust the system scanning parameters. The model calculates the patient entrance dose from an energy spectrum generated using the input parameters. The effective dose for a paediatric sized patient is then calculated using a Monte Carlo simulation. The dynamic model allows for variation of the scan parameters and direct observation of the expected dose levels for specific examinations. Filtration is a well-known technique for reducing radiation dose, where a filter material is placed in the path of the X-ray beam to reduce patient exposure to radiation. The dynamic model was used to design a new filtration technique for the paediatric settings on the Statscan.en_ZA
dc.identifier.apacitationPerks, T. D. (2014). <i>Reducing radiation dose for a linear slot scanning digital X-ray machine using a filtration technique</i>. (Thesis). University of Cape Town ,Faculty of Health Sciences ,Division of Biomedical Engineering. Retrieved from http://hdl.handle.net/11427/13968en_ZA
dc.identifier.chicagocitationPerks, Timothy David. <i>"Reducing radiation dose for a linear slot scanning digital X-ray machine using a filtration technique."</i> Thesis., University of Cape Town ,Faculty of Health Sciences ,Division of Biomedical Engineering, 2014. http://hdl.handle.net/11427/13968en_ZA
dc.identifier.citationPerks, T. 2014. Reducing radiation dose for a linear slot scanning digital X-ray machine using a filtration technique. University of Cape Town.en_ZA
dc.identifier.ris TY - Thesis / Dissertation AU - Perks, Timothy David AB - This study describes the development of a filtration technique applied to the Lodox Statscan linear slotscanning digital X-ray system to reduce radiation dose to paediatric patients whilst preserving diagnostic image quality. The Statscan is an FDA approved, commercially available digital X-ray system commonly used for trauma and emergency patients. The Statscan provides significantly lower radiation dose to patients than conventional Xray systems for comparable studies without loss of image quality. This is particularly beneficial in paediatric radiology, where the risks associated with ionizing radiation are much higher. A static dose prediction model for the Statscan which was previously developed at the University of Cape Town has been adapted to create a dynamic dose prediction model which allows the user to adjust the system scanning parameters. The model calculates the patient entrance dose from an energy spectrum generated using the input parameters. The effective dose for a paediatric sized patient is then calculated using a Monte Carlo simulation. The dynamic model allows for variation of the scan parameters and direct observation of the expected dose levels for specific examinations. Filtration is a well-known technique for reducing radiation dose, where a filter material is placed in the path of the X-ray beam to reduce patient exposure to radiation. The dynamic model was used to design a new filtration technique for the paediatric settings on the Statscan. DA - 2014 DB - OpenUCT DP - University of Cape Town LK - https://open.uct.ac.za PB - University of Cape Town PY - 2014 T1 - Reducing radiation dose for a linear slot scanning digital X-ray machine using a filtration technique TI - Reducing radiation dose for a linear slot scanning digital X-ray machine using a filtration technique UR - http://hdl.handle.net/11427/13968 ER - en_ZA
dc.identifier.urihttp://hdl.handle.net/11427/13968
dc.identifier.vancouvercitationPerks TD. Reducing radiation dose for a linear slot scanning digital X-ray machine using a filtration technique. [Thesis]. University of Cape Town ,Faculty of Health Sciences ,Division of Biomedical Engineering, 2014 [cited yyyy month dd]. Available from: http://hdl.handle.net/11427/13968en_ZA
dc.language.isoengen_ZA
dc.publisher.departmentDivision of Biomedical Engineeringen_ZA
dc.publisher.facultyFaculty of Health Sciencesen_ZA
dc.publisher.institutionUniversity of Cape Town
dc.subject.otherBiomedical Engineeringen_ZA
dc.titleReducing radiation dose for a linear slot scanning digital X-ray machine using a filtration techniqueen_ZA
dc.typeMaster Thesis
dc.type.qualificationlevelMasters
dc.type.qualificationnameMSc (Med)en_ZA
uct.type.filetypeText
uct.type.filetypeImage
uct.type.publicationResearchen_ZA
uct.type.resourceThesisen_ZA