Control and monitoring of an intra-aortic balloon pump for cardiac assist

Master Thesis

1971

Permanent link to this Item
Authors
Journal Title
Link to Journal
Journal ISSN
Volume Title
Publisher
Publisher

University of Cape Town

License
Series
Abstract
An anatomical and physiological background to the heart as a pump, followed by a description of the causes and results of cardiogenic shock is given. A survey of the achievements of other researchers in this field is included. The design of a system for controlling and monitoring the Intra-Aortic Balloon Pump is described. A narrow band-pass filter is used to reduce noise and limit the input to the detector to the spectrum within which the R-wave of an ECG signal lies. A safety circuit to prevent the heart pumping against an inflated balloon is incorporated. History of the 8 secs preceding any instant is available. A description of the medical tests carried out so far is given.
Description

Reference:

Collections